OnSite - establishing a new standard for Medical Waste Autoclaves and Medical Waste Treatment
The operational programming for The VariClave® has been thoughtfully developed and thoroughly tested. The
proprietary VariClave® Three Phase Process and Automatic Variable Parametric Settings ensure efficient and
efficacious treatment of your Regulated Medical Waste.
•
The installed system programming includes: cycle programming and a recordkeeping database of cycle
history, machine usage, operator training and alarm occurrence.
•
The cycle programming capability provides the user with a selection of cycle types. The operating
characteristics of each cycle type have been individually tested and validated for efficacy to suit each material
type and its containerization method.
•
There are three wet steam cycles standard and one dry heat cycle available. The wet steam cycles have three
phases:
•
Phase 1 is for heat-up, the treatment of airborne pathogens and air removal. This cycle can be repeated
up to five times to ensure the complete removal of air from the chamber and process material.
•
Phase 2 is for the sterilization of the material. During this phase, the chamber is maintained at the
pressure and temperature set points for a period to heat the process material to the “kill” temperature
for sterilization.
•
Phase 3 employs the vacuum pump to dry and cool the process material and a directional spray
manifold to cool the bin for safe removal.
•
The program is designed to automatically vary the process parameters based on the cycle type selected and the
weight of the material to be processed.
•
The programming also includes a searchable database. Information regarding usage, process cycles,
maintenance, operator training and alarms is recorded in the database. This information is maintained for a
period of seven years in a compressed format on the systems 4GB hard drive. After seven years of data logging,
the oldest data is automatically deleted from the database.
•
A writable CD ROM drive is also included along with a file export system for removable storage of data.